REVERSE OSMOSIS
- Raw water pump – To force raw water through Sand filter, Activated Carbon Filter, Pre RO Micron filter and ANTISCALANT DOSING SYSTEM.
- Sand Filter – To Filter out suspended particulate matter greater than 25 Microns in size.
- Activated Carbon Filter – To remove smell, odour and other organic contaminants and to protect RO membrane from free oxidants.
- Micron Filter – To remove suspended particulate matter up to 5 Microns in size.
- RO Unit – To remove suspended and dissolved solids including micro organisms and to produce treated water confirming to AAMI standards for Haemodialysis
MICRON FILTRATION & UV STERILIZATION
- Treated water transfer pump – To pump the water through Micron cartridge filter rated 1 & 0.2 Microns and through UV Steriliser.
- 1 Micron Filter – To Filter out particles up to 1 Micron in Size and to protect 0.2 Micron rated cartridge from frequent choking.
- 0.2 Micron Filter – To filter out bacteria and pyrogen
- UV Steriliser – For sterilization of water.
Capacity of RO system in Ltrs per Hour
Single Stage | Double Stage |
250 LPH | 500 + 250 LPH |
500 LPH | 1000 + 500 LPH |
1000 LPH | 2000 + 1000 LPH |
2000 LPH | 3000 + 2000 LPH |
Post Treatment System | UF System as per required Capacity |
Portable RO system 50 LPH | Portable RO system 100 + 50 LPH |