Call(Office): +91 9590007591
Mon - Fri: 9.00am - 7.00pm

REVERSE OSMOSIS SYSTEM FOR HEMODIALYSIS

REVERSE OSMOSIS

  1. Raw water pump – To force raw water through Sand filter, Activated Carbon Filter, Pre RO Micron filter and ANTISCALANT DOSING SYSTEM.
  2. Sand Filter – To Filter out suspended particulate matter greater than 25 Microns in size.
  3. Activated Carbon Filter – To remove smell, odour and other organic contaminants and to protect RO membrane from free oxidants.
  4. Micron Filter – To remove suspended particulate matter up to 5 Microns in size.
  5. RO Unit – To remove suspended and dissolved solids including micro organisms and to produce treated water confirming to AAMI standards for Haemodialysis.

Salient Features:

  1. Compact Design.
  2. Custom built design to suit varied feed water condition.
  3. Fully automated using simple logic controller.
  4. Treated water quality confirming to AAMI standards.
  5. Optimum RO system recovery to reduce waste water quantity to less than 35%.
  6. System designed for rugged operation using best quality of components available in the industry.
  7. Offering timely & quality after sales service with dedicated and well experienced service team.

Micron Filtration & UV Sterilization

  1. Treated water transfer pump – To pump the water through Micron cartridge filter rated 1 & 0.2 Microns and through UV Steriliser.
  2. 1 Micron Filter – To Filter out particles up to 1 Micron in Size and to protect 0.2 Micron rated cartridge from frequent choking.
  3. 2 Micron Filter – To filter out bacteria and pyrogen
  4. UV Steriliser – For sterilization of water.

Capacity of RO system in Ltrs per Hour

Single Stage Double Stage
250 LPH 500 + 250 LPH
500 LPH 1000 + 500 LPH
1000 LPH 2000 + 1000 LPH
2000 LPH 3000 + 2000 LPH
Post Treatment System UF System as per required Capacity
Portable RO system 50 LPH Portable RO system 100 + 50 LPH